V214.1 (V-113.1 in journal) Exploring the role of an implantable doppler probe as a blood flow-monitoring device in kidney transplant: A feasibility randomized controlled trial

Shahzar Malik, United Kingdom

Trainee Surgery
Southwest Transplant Centre
NHS

Abstract

Exploring the role of an implantable doppler probe as a blood flow-monitoring device in kidney transplant: A feasibility randomized controlled trial

Shahzar Malik1.

1Southwest Transplant Centre, University Hospitals Plymouth, National Health Service Trust, Plymouth, United Kingdom

Objectives: Vascular thrombosis is a disastrous postoperative complication resulting in the loss of 3.5% to 5.7% of all kidney transplants. The use of blood flow-sensing technology in the early postoperative period may help in the early identification of vascular complications crucial to reducing graft loss. This study aimed to assess the feasibility of an implantable Doppler probe as a blood flow-monitoring device in kidney transplant recipients and to evaluate its usefulness in preventing early graft loss.
Materials and Methods: This 2-arm feasibility randomized controlled trial compared the demographical characteristics and surgical outcomes of kidney transplant recipients who received implantable Doppler probe monitoring (intervention group; n = 30) with those who had standard clinical care (control group; n = 30). Surgical outcomes compared between the groups included the number of early vascular complications identified, the number of departmental ultrasonography scans requested in the first 72 hours postoperatively, and month 3 graft loss.
Results: Both groups were similar in demographic characteristics. In the intervention group versus the control group, fewer ultrasonography scans were requested in the first 24 hours postoperatively (56% vs 91%) and lower graft loss (0% vs 6.6%) was recorded. The results addressed uncertainties around the feasibility study’s research methods and required resources for a future pragmatic trial.
Conclusions: An implantable Doppler probe may be a beneficial adjunct for graft monitoring after kidney transplants. This feasibility study provided the necessary preliminary information and filled initial gaps in the evidence that can inform future research. The prespecified progression criteria of the study were fulfilled. The study template used can be transferable to other transplant centers across the United Kingdom. A pragmatic large-scale randomized controlled trial is warranted to evaluate the effectiveness of implantable Doppler probes in clinical practice.

References:

[1] Blood flow-sensing technology
[2] Graft loss
[3] Renal transplant